Foerster SIGMATEST

Portable electrical conductivity meter for non-ferromagnetic metals using eddy currents. Meets Boeing (BAC 5651) and Airbus requirements.

Five frequencies
50 measurements/s
Lift-off 0.75 mm
Temperature compensation
SIGMATEST
SIGMATEST
SIGMATEST
1 / 3
SIGMATEST

Key features

  • Five test frequencies ranging from 60 to 960 kHz for different material thicknesses.
  • Up to 50 measurements per second in Touch or Continuous mode.
  • Normalizes values to 20 °C using an internal or external temperature sensor.
  • Measures through paint and coatings with a lift-off of up to 0.75 mm.
  • Automatically recognizes the probe and applies its calibration curve.
  • SD card storage and Ethernet interface for system integration.
Range0.5 to 65 MS/m (1 to 112% IACS)
Accuracy±0.7% of the measured value
Frequencies60 to 960 kHz

Display

TypeLCD
Resolution240 × 320 px

Measurement

Measurement range0.5 to 65 MS/m (1 to 112% IACS)
Accuracy±0.7% of the measured value, at 60 kHz with a 14-mm probe
Reading resolution±0.1% of the measured value
UnitsMS/m, % IACS
Measurement speedUp to 50 measurements/s
Lift-off compensationUp to 0.75 mm (0.03 in), automatic
Temperature compensationAutomatic, normalized to 20 °C; sensor integrated into the probe or optional external sensor
CalibrationCalibration data stored in each probe, automatic curve selection

Evaluation

FunctionsCorrection factors for concave and convex surfaces; application-specific temperature coefficients; multilingual interface

Eddy currents

Test frequency60, 120, 240, 480, and 960 kHz
Compatible Probes5 mm, 8 mm, and 14 mm print head widths, automatic detection

Data

StorageSD card, virtually unlimited capacity

Connectivity

Ports10-pin LEMO connector, SD card slot, 100 Mbit/s RJ45 Ethernet

Power

Battery Life4 hours per battery set
Charger5 V DC / 3,000 mA, adaptable to any country’s voltage

Environmental

Operating Temperature0 to 40 °C (32 to 104 °F)
Humidity5 to 85%

Physical

Dimensions211 × 102 × 40 mm
Weight620 g

Standards

ComplianceDIN 50994, DIN EN 2004-1:1993-09, DIN EN 2004-7:2017-10, ASTM E 1004, MIL STD 1537C, AITM 6 to 6008; Boeing (BAC 5651) and Airbus requirements

Probes

SIGMATEST 5 mm ProbeHead width 5 mm
SIGMATEST 8 mm ProbeHead width 8 mm
SIGMATEST 14 mm ProbeHead width 14 mm

Optional Accessories

External Temperature SensorFor when the workpiece and probe temperatures differ
Coin Testing StationStation for reproducible positioning in conductivity measurements
SIGMATEST Power KitExtends operating time during mobile measurements
Conductivity Reference StandardsFor verification and adjustment of conductivity measurements
Aerospace